USP Class VI describes the result for a specimen tested under USP 〈88〉. ISO 10993-1 sets out how to evaluate biological safety for a device in its risk management process. For non-implantable silicone parts, ask what was tested and map the report to the finished device and its intended contact.
Key Sourcing Takeaways
- Check the grade, specimen, extraction conditions and report revision.
- Describe direct or indirect contact, its duration and repeat use.
- Record any differences between the tested specimen and the finished device.
Four-Stage Lot Inspection & Raw-to-Shipment Traceability
- 01
Material grade & batch COA
- 02
Mixing & molding records
- 03
2D optical FAI records
- 04
SGS/TÜV finished-part scope

What each reference covers
| Reference | Scope | What to ask for |
|---|---|---|
| USP Class VI under USP 〈88〉 | In vivo biological reactivity classification of the test material or specimen under specified conditions | The chapter version, specimen and formulation, extraction media and conditions, and individual test results |
| ISO 10993-1:2025 | Biological safety evaluation of a medical device within risk management | Intended use, exposure details, materials and manufacturing information, relevant existing evidence and the rationale for any additional testing |
USP General Chapter 〈88〉 describes systemic injection, intracutaneous and implantation tests. The implantation test in the chapter concerns the test method. The device's intended use determines its contact category.
ISO 10993-1:2025 sets requirements and principles for biological safety evaluation. For US submissions, check the applicable edition, extent of recognition and transition terms in the FDA recognized standards database.
Compare the specimen with the final finished device
A supplier's test report may describe a molded plaque in one formulation. Check whether the production part uses the same grade, colorants and cure. Include relevant process residues, adhesives, packaging and sterilization in the comparison. A changed formulation or process calls for a documented review of whether the evidence still applies.
For a wearable sensor mount, identify the skin-facing silicone, adhesive and other layers as used. For a mask interface or nebulizer connector, describe any direct contact and any indirect exposure through the gas or fluid path. Include contact duration and repeated use. The device manufacturer uses these details to justify the evidence and address gaps for the specific finished device. The FDA guidance on ISO 10993-1 explains the role of the finished device and representative test articles in US submissions.
What to put on the RFQ
Send the drawing revision, intended use, contact route and duration, expected production compound and processing, and the tests or documents your quality team requires. Ask the supplier to identify the test specimen and explicitly state which materials and processes the provided report covers. Agree on how changes will be reviewed before approving production samples.