USP Class VI describes the result for a specimen tested under USP 〈88〉. ISO 10993-1 sets out how to evaluate biological safety for a device in its risk management process. For non-implantable silicone parts, ask what was tested and map the report to the finished device and its intended contact.

Molding Process & DFM DesignUpdated

3 min readCompiled by Youbang Silicone Engineering Team
Direct Engineering Answer

USP Class VI describes the result for a specimen tested under USP 〈88〉. ISO 10993-1 sets out how to evaluate biological safety for a device in its risk management process. For non-implantable silicone parts, ask what was tested and map the report to the finished device and its intended contact.

Key Sourcing Takeaways

  • Check the grade, specimen, extraction conditions and report revision.
  • Describe direct or indirect contact, its duration and repeat use.
  • Record any differences between the tested specimen and the finished device.

Four-Stage Lot Inspection & Raw-to-Shipment Traceability

  1. 01

    Material grade & batch COA

  2. 02

    Mixing & molding records

  3. 03

    2D optical FAI records

  4. 04

    SGS/TÜV finished-part scope

Engineering process control benchmark · final tolerances confirmed per project CAD
Compression-molding operation detail — Youbang Silicone
Compression-molding operation detail

What each reference covers

Reference Scope What to ask for
USP Class VI under USP 〈88〉 In vivo biological reactivity classification of the test material or specimen under specified conditions The chapter version, specimen and formulation, extraction media and conditions, and individual test results
ISO 10993-1:2025 Biological safety evaluation of a medical device within risk management Intended use, exposure details, materials and manufacturing information, relevant existing evidence and the rationale for any additional testing

USP General Chapter 〈88〉 describes systemic injection, intracutaneous and implantation tests. The implantation test in the chapter concerns the test method. The device's intended use determines its contact category.

ISO 10993-1:2025 sets requirements and principles for biological safety evaluation. For US submissions, check the applicable edition, extent of recognition and transition terms in the FDA recognized standards database.

Compare the specimen with the final finished device

A supplier's test report may describe a molded plaque in one formulation. Check whether the production part uses the same grade, colorants and cure. Include relevant process residues, adhesives, packaging and sterilization in the comparison. A changed formulation or process calls for a documented review of whether the evidence still applies.

For a wearable sensor mount, identify the skin-facing silicone, adhesive and other layers as used. For a mask interface or nebulizer connector, describe any direct contact and any indirect exposure through the gas or fluid path. Include contact duration and repeated use. The device manufacturer uses these details to justify the evidence and address gaps for the specific finished device. The FDA guidance on ISO 10993-1 explains the role of the finished device and representative test articles in US submissions.

What to put on the RFQ

Send the drawing revision, intended use, contact route and duration, expected production compound and processing, and the tests or documents your quality team requires. Ask the supplier to identify the test specimen and explicitly state which materials and processes the provided report covers. Agree on how changes will be reviewed before approving production samples.

Turn the selection into a reviewable project brief

Sourcing & Tooling FAQ

Key Answers Before Tooling

Covering CAD readiness, 100 pc trial & 1,000 pc / USD 2,000 standard MOQ policy, free 3D fit prototypes vs. T1 silicone samples, NDA protection, and mold ownership.

Have a part-specific question?

Send your drawing or sample photo directly to Youbang’s engineering desk; we send an initial response within 24 hours.

What should a Class VI report identify?

Check the grade, specimen, extraction conditions and report revision.

What contact information does the device team need?

Describe direct or indirect contact, its duration and repeat use.

What if the report used a molded plaque?

Record any differences between the tested specimen and the finished device.

Can the same material report support a new color?

Record changes to colorants and formulation. The device team reviews whether existing evidence applies and whether additional evaluation is needed.

B2B Factory Direct · initial response within 24 hours

Review the component drawing

Describe direct or indirect contact, its duration and repeat use.

✓ Trial from 100 pcs · Standard 1,000 pcs or USD 2,000✓ Free 3D-printed fit prototypes before steel tooling✓ Paid molds 100% owned by customer

Send Your Project Brief or CAD

First reply within 24 hours · Trial orders from 100 pcs

For confidential drawings, sign the mutual NDA before uploading the files.

Only rights holders or their authorized representatives may upload files. We keep inquiries and attachments to prepare quotes and follow up on your project.

Prefer email or files > 20 MB? Send directly to inquiry@ybsilicone.com · Privacy Policy